5 Parameter Patient Monitor 12.1″
5 Multi-Parameter Patient Monitor 12.1″ Supplied by Ambulance Products India
Specs:-
Quality Standards and Classification
CE, IS013485
SFDA: Class II b
Anti-electroshock degree:
Class I equipment
(internal power supply)
TEMP/SpO 2 /NIBP: BF
ECG/Resp: CFApplication Range
Adult / Paediatric / Neonatal / Medicine / Surgery /
Operating Room / lCU / CCU
Display
12.1″ real color TFT screenEnvironment
Operating environment :
Temperature: 0 ~40 C
Humidity: 15%~80%
Pressure: 86.0 kPa~106.0kPa
Transport and Storage environment :
Temperature: -20 ~55 C
Humidity: 10%~90%
Pressure: 86.0 kPa~106.0kPa
CE, IS013485
SFDA: Class II b
Anti-electroshock degree:
Class I equipment
(internal power supply)
TEMP/SpO 2 /NIBP: BF
ECG/Resp: CFApplication Range
Adult / Paediatric / Neonatal / Medicine / Surgery /
Operating Room / lCU / CCU
Display
12.1″ real color TFT screenEnvironment
Operating environment :
Temperature: 0 ~40 C
Humidity: 15%~80%
Pressure: 86.0 kPa~106.0kPa
Transport and Storage environment :
Temperature: -20 ~55 C
Humidity: 10%~90%
Pressure: 86.0 kPa~106.0kPa
Power Requirements
AC: 100 – 240V, 50Hz/60Hz
DC: Built-in rechargeable battery
Battery: 11.1V 24wh Li-ion battery
2h operating after full charge
5min operating after low battery alarm
Dimension and Weight
Equipment:
340mm x 245mm x134 mm (about);
2.798kg (about)
Package:
380mm x350 mm x 300mm; 4.598 kg (about)Date Storage
Trend diagram/table: 7x24h
N I BP review: 400 events
Wave review: 100min
Alarm review: 100 alarm events
Support drug concentration titration analysisECG
5 Leads: RA, LA, LL, RL, V
Lead mode: I, II, III, AVR, AVL, AVF, V
Gain: x 1.25mm/mV,2.5mm/mV,
5.0mm/mV, 10mm/mV,
20mm/mV (error≥ 5%)
Sweep speed: 6.25mm/s,12.5mm/s,25mm/s,
50mm/s (error ≥ ± 10%)
Bandwidth: Surgery 1~25 Hz
Monitor 0.5 ~40 Hz
Diagnostic 0.05 ~130 Hz
Sensitivity: >200 µ V P-P
CMRR: ≥112dB
Differential Input Impedance: >5M
ohm Electrode Offset Potential: ±300mV
Polarization Voltage: ≥ ± 500mV
Input Impedance: ≥ 5 M
Calibration Voltage: 1 mV ± 5%
Equipment:
340mm x 245mm x134 mm (about);
2.798kg (about)
Package:
380mm x350 mm x 300mm; 4.598 kg (about)Date Storage
Trend diagram/table: 7x24h
N I BP review: 400 events
Wave review: 100min
Alarm review: 100 alarm events
Support drug concentration titration analysisECG
5 Leads: RA, LA, LL, RL, V
Lead mode: I, II, III, AVR, AVL, AVF, V
Gain: x 1.25mm/mV,2.5mm/mV,
5.0mm/mV, 10mm/mV,
20mm/mV (error≥ 5%)
Sweep speed: 6.25mm/s,12.5mm/s,25mm/s,
50mm/s (error ≥ ± 10%)
Bandwidth: Surgery 1~25 Hz
Monitor 0.5 ~40 Hz
Diagnostic 0.05 ~130 Hz
Sensitivity: >200 µ V P-P
CMRR: ≥112dB
Differential Input Impedance: >5M
ohm Electrode Offset Potential: ±300mV
Polarization Voltage: ≥ ± 500mV
Input Impedance: ≥ 5 M
Calibration Voltage: 1 mV ± 5%
ST Segment
ST Segment Range: -2.0mV ~ +2.0mV
Accuracy: 0.02mVNIBP
Method: Oscillometry
Measure mode: Manual, Auto, STAT
Measure interval in auto mode:
1~5/10/15/30/60/90/120/180/240/480min
Measure period in STAT mode: 5min
Pulse rate range: 40 ~240bpm
Measure & alarm range: Adult
SYS 40~280mmHg
DIA 10~220mmHg
MEAN 20~240mmHg
Paediatric
SYS 40~220mmHg
DIA 10~160mmHg
MEAN 20~170mmHg
Neonatal
SYS 40~135mmHg
DIA 10~100mmHg
MEAN 20~110mmHg
Static pressure measurement
Range: 0 ~ 300mmHg
Accuracy: ± 3mmHg
Resolution: 1 mmHg Accuracy:
Max. Mean Error: ± 5mmHg
Max. Standard Deviation: 8mmHg
Overpressure protection:
Adult 300mmHg
Paediatric 240mmHg
Neonatal 150mmHg
ST Segment Range: -2.0mV ~ +2.0mV
Accuracy: 0.02mVNIBP
Method: Oscillometry
Measure mode: Manual, Auto, STAT
Measure interval in auto mode:
1~5/10/15/30/60/90/120/180/240/480min
Measure period in STAT mode: 5min
Pulse rate range: 40 ~240bpm
Measure & alarm range: Adult
SYS 40~280mmHg
DIA 10~220mmHg
MEAN 20~240mmHg
Paediatric
SYS 40~220mmHg
DIA 10~160mmHg
MEAN 20~170mmHg
Neonatal
SYS 40~135mmHg
DIA 10~100mmHg
MEAN 20~110mmHg
Static pressure measurement
Range: 0 ~ 300mmHg
Accuracy: ± 3mmHg
Resolution: 1 mmHg Accuracy:
Max. Mean Error: ± 5mmHg
Max. Standard Deviation: 8mmHg
Overpressure protection:
Adult 300mmHg
Paediatric 240mmHg
Neonatal 150mmHg
Sp02
Measurement range: 0~100%
Resolution: 1%
Accuracy: 70%~100% ± 2%
40%~69% ± 3%Pulse Rate
Measure/alarm range: 20 ~300bpm
Resolution: 1 bpm
Accuracy: ± 3bpmRESP
Method: impedance between RA-LL
Resp Impedance range: 0.3 – 3 ?
Base Impedance range: 200 Q – 4000 ?
Bandwidth: 0.3~2.5Hz
Gain: x 0.25, x 0.5, x 1, x 2, x 4
Resp rate:
Measurement Range
Adult 0~120BrPM
Neonatal/Paediatric 0~150BrPM
Resolution: 1BrPM
Accuracy: 7~150BrPM (± 2BrPM)
0 – 6BrPM (unspecified)
Apnea Alarm: 10 ~40s
Measurement range: 0~100%
Resolution: 1%
Accuracy: 70%~100% ± 2%
40%~69% ± 3%Pulse Rate
Measure/alarm range: 20 ~300bpm
Resolution: 1 bpm
Accuracy: ± 3bpmRESP
Method: impedance between RA-LL
Resp Impedance range: 0.3 – 3 ?
Base Impedance range: 200 Q – 4000 ?
Bandwidth: 0.3~2.5Hz
Gain: x 0.25, x 0.5, x 1, x 2, x 4
Resp rate:
Measurement Range
Adult 0~120BrPM
Neonatal/Paediatric 0~150BrPM
Resolution: 1BrPM
Accuracy: 7~150BrPM (± 2BrPM)
0 – 6BrPM (unspecified)
Apnea Alarm: 10 ~40s
TEMP
Channel 2
Measure & alarm range: 0~50 °C
Resolution: 0.1 °C
Accuracy (no sensor): ±0.1 C (0~50 °C)
Standard accessories
• NIBP cuff & tube
• ECG cable & electrodes
• SpO2 sensor
• TEMP probe
• Lithium-ion battery
• Power cable
• Operator’s manual
We are Exporters and Suppliers of 5 Multi-Parameter Patient Monitor 12.1″ at competitive prices from India.
• NIBP cuff & tube
• ECG cable & electrodes
• SpO2 sensor
• TEMP probe
• Lithium-ion battery
• Power cable
• Operator’s manual
We are Exporters and Suppliers of 5 Multi-Parameter Patient Monitor 12.1″ at competitive prices from India.
For More Information, visit us at www.descomedicalindia.com
Model No
IEPM 110 5 Parameter Patient Monitor 12.1″
Reviews
There are no reviews yet.